YSI ProDIGITAL ODO/T PROBE ASSEMBLY Cable 4 METER



The ProDIGITAL ODO/T probe and cable assembly features integral optical dissolved oxygen and temperature sensors. Cable lengths are available up to 100 meters.

The ODO/CT assembly is compatible with the ProSoloProSwap, and ProDSS handhelds (sold separately). Pre-assembled ProSolo ODO/T kits are available.

 

The ODO/T probe assembly pairs with the ProSolo Handheld to form a complete sampling system. Please visit the ProSolo ODO/T Kits page for fully assembled systems.

Featuring integral, digital sensors, the ODO/T is plug-and-play which minimizes time spent on instrument setup. ProDIGITAL technology allows the handheld to automatically recognize connected sensors, so there is no need to tell the instrument what is connected.

Optical Dissolved Oxygen (ODO) technology is non-consumptive, meaning oxygen is not consumed during the measurement. Flow dependence is non-existent and no stirring is required, resulting in the most accurate DO measurements available from a probe. The ODO sensor is not susceptible to interferences from other gases, drifts less and holds calibration longer than membrane covered sensors. Maintenance is a breeze with a simple, threaded optical cap that is good for at least two years.

The ProDIGITAL ODO/T probe and cable assembly features include:

  • Integral optical dissolved oxygen, and temperature sensors that are auto-recognized by ProDIGITAL handhelds, reducing instrument set-up time
  • Rugged Optical DO sensor cap with a 2-year warranty is pre-installed on new assemblies
  • Military Spec (MS) connector and a weighted, stainless steel probe guard
  • Fits in a 1 inch well with cable lengths up to 100 meters
  • 2-year warranty on cable, sensors, and ODO cap

Specifications

SKU: 627200-4

ProDIGITAL ODO Cable The ProDIGITAL ODO/T probe and cable assembly features integral optical dissolved oxygen and temperature sensors:
Accuracy
Temperature - ±0.2 °C

Dissolved Oxygen - 0 to 200%: ±1% of reading or 1% saturation, whichever is greater 200 to 500%: ±8% of reading 0 to 20 mg/L: ±0.1 mg/L or 1% of reading, whichever is greater 20 to 50 mg/L: ±8% of reading

Measurement Method / Channels Temperature - Thermistor; Dissolved Oxygen - Optical luminescence - lifetime method
Measurement Range
Temperature - -5 to 70 °C (temperature compensation range for DO mg/L measurement: -5 to 50 °C )

Dissolved Oxygen - 0 to 500%, 0 to 50 mg/L

Parameters Measured
Dissolved Oxygen, Temperature

Size
Width - 2.46 cm (0.97 in.)

Unit of Measure Temperature - °C, °F, K; Dissolved Oxygen - % saturation, % saturation local, mg/L, ppm
Warranty
Warranty 2-year probe and cable assembly; 2-year extended warranty ODO sensor cap